Quick answer. The Mindray BeneVision N-series (N1, N12, N15, N17, N19, N22) is Mindray's current-generation modular bedside platform, replacing the earlier BeneView T-series. Two architectural facts drive accessory selection: bedside / ICU models use Mindray's MPM acquisition module while the N1 transport monitor uses a built-in parameter set — which determines connector pinout; and SpO2 technology is selectable per unit (Mindray-native, Masimo SET, or Nellcor OxiMax), with the connector and sensor family changing with that choice. Jump to the complete directly-linked accessory list by family.
This is a selection hub, not a buying guide. It links MedLinket accessories matched to the BeneVision N-series, organised by parameter and by clinical scenario. For the full catalogue, see the Mindray compatible accessories collection.
BeneVision N-series models covered
| Model | Display | Typical clinical unit | Parameter architecture |
|---|---|---|---|
| N1 | 8" transport-capable | Transport, ED triage, sub-acute | Built-in SpO2 / ECG / NIBP plus a slot for limited modules |
| N12 | 12" | Step-down, sub-acute, pediatric ward | MPM module-based |
| N15 | 15" | General ICU, CCU | MPM module-based, full module rack |
| N17 | 17" | Adult ICU, surgical ICU | MPM module-based, full module rack |
| N19 | 19" | Cardiothoracic ICU, transplant | MPM, dual-screen capable |
| N22 | 22" | Cardiac surgery OR, complex ICU | MPM, dual-screen, advanced hemodynamic |
How to identify your BeneVision SpO2 technology
Three ways to confirm which SpO2 technology is installed:
- Look at the SpO2 cable end on the MPM module: Mindray-native is a 7-pin Mindray connector; Masimo SET is the LNCS connector; Nellcor OxiMax is the 9-pin connector.
- Check the MPM module label — Mindray labels the SpO2 technology variant on the module casing.
- Check the service menu: System Information → Module Status → SpO2 → Technology.
A wrong-technology sensor produces a "Sensor not recognized" alarm — no harm to patient or monitor. For visual references, see what "compatible" actually means.
SpO2 accessories by host technology — the rule
- Mindray-native SpO2 — used most often on the N1 and on units sold in China and select Asia-Pacific markets, with the 7-pin Mindray connector. MedLinket Mindray-native compatible sensors include reusable adult / pediatric finger clips, an adult soft-tip sensor, a neonate wrap and a multi-site Y sensor, plus a 7-pin extension cable and disposable adhesive sensors.
- Masimo SET — the SET algorithm runs on the Mindray-licensed Masimo board in the MPM module. MedLinket LNCS-compatible cables transmit the optical waveform and do not replicate any portion of the SET algorithm. See the Masimo LNCS / RD / rainbow compatible sensors hub.
- Nellcor OxiMax — MedLinket offers the DS-100A-style reusable adult finger clip and MAX-A-style disposable sensor. For bench data see the Nellcor sensors on Mindray BeneVision monitors test.
Compatible accessories by family — direct links
Every accessory below is matched to the BeneVision N-series. Confirm each unit's SpO2 technology (Mindray-native / Masimo SET / Nellcor) before ordering sensors, and confirm the N1 6-pin vs MPM 12-pin connector for ECG. For anything not listed, browse the full Mindray compatible accessories collection.
SpO2 sensors by host technology
All Mindray SpO2| Product | Patient / type | Host |
|---|---|---|
| 512FLH adult finger clip | Adult, reusable | Mindray-native |
| Adult silicone soft | Adult, reusable soft-tip | Mindray-native |
| Adult ear clip | Adult, reusable | Mindray-native |
| Neonate silicone wrap (115-020887-00) | Neonate, reusable wrap | Mindray-native |
| Multi-site Y sensor | All patient, multi-site | Mindray-native |
| Masimo LNCS (1864) | Adult finger, reusable | Masimo SET |
| DS-100A compatible | Adult finger, reusable | Nellcor |
| MAX-A compatible | Adult, disposable adhesive | Nellcor |
SpO2 adapter / extension cables
All SpO2 cables| OEM cross-reference | Configuration |
|---|---|
| 115-020768-00 compatible | Mindray SpO2 adapter cable |
| 115-023135-00 compatible | Mindray SpO2 adapter cable |
| 115-023136-00 compatible | Mindray SpO2 cable |
| 0010-20-42710 compatible | Mindray/Datascope SpO2 adapter |
| 0010-20-42594 compatible | Mindray/Datascope SpO2 adapter |
| 0010-20-43075 compatible | Mindray/Datascope SpO2 adapter |
ECG cables & leadwires 12-pin MPM / 6-pin N1
All ECG| OEM cross-reference | Type |
|---|---|
| 0010-30-42782 compatible | ECG trunk cable |
| 0010-30-42719 compatible | ECG trunk cable |
| 0010-30-12377 compatible | ECG trunk cable |
| 0010-30-42723 compatible | ECG trunk cable |
| Neonatal 3-lead DIN trunk | Neonatal trunk cable |
| EA6232B compatible | 3-lead, pinch/snap (IEC) |
| EA6232A compatible | 3-lead, pinch/grabber (IEC) |
| EA6252B compatible | 5-lead, pinch/snap (IEC) |
| EA6252A compatible | 5-lead, pinch/grabber (IEC) |
| 040-000960-00 compatible | 3-lead, grabber (AHA) |
| 040-000961-00 compatible | 5-lead, snap (AHA) |
| Mindray 5-lead leadwires (AHA grabber) | Leadwire set, molded |
| EY6501A compatible | Telemetry leadwire, 5-ld clip (IEC) |
| EY6501B compatible | Telemetry leadwire, 5-ld snap (IEC) |
| 115-004867-00 compatible | Telemetry leadwire |
ECG electrodes universal
All electrodes| Product | Patient / type |
|---|---|
| Adult solid-gel, 4.0mm snap | Adult |
| Pediatric electrodes | Pediatric |
| Disposable ECG electrodes | General |
Electrodes are interchangeable across Mindray BeneVision and BeneView and most other OEM platforms — buy by patient type and clinical scenario, not by monitor model.
NIBP cuffs & hoses
All NIBP| Product | Size / type |
|---|---|
| Reusable pediatric cuff | Reusable (fit your Mindray connector) |
| Reusable adult cuff | Reusable (fit your Mindray connector) |
| Reusable adult large cuff | Reusable (fit your Mindray connector) |
| Reusable neonatal cuff | Reusable (fit your Mindray connector) |
| Disposable neonate size 1 (3–6 cm) | Disposable, sizes 1–5 available |
| NIBP hose 6200-30-09688 | Adult NIBP hose |
| NIBP hose 6200-30-11560 | Neonatal luer-slip hose |
| 989803136941 | Neonate NIBP hose |
IBP cables 12-pin Mindray
All IBP cables| OEM cross-reference | Transducer interface |
|---|---|
| 0010-21-12179 compatible | Edwards |
| 650-206M compatible | Utah |
| B. Braun (12 ft) | B. Braun |
| Argon (12 ft) | Argon / Maxxim |
| 001C-30-70757 compatible | BD |
| 001C-30-70759 compatible | Abbott |
| 040-001029-00 extension | IBP extension (double-end) |
Pair with single-use disposable IBP transducers by connector standard. See the Edwards IBP transducer cross-brand test.
Temperature probes & adapters YSI-400
All temperature| OEM cross-reference | Type |
|---|---|
| 0011-30-37393 adult skin | Reusable, YSI-400 |
| MR401B adult eso/rectal | Reusable, YSI-400 |
| Pediatric skin probe | Reusable, YSI-400 |
| Pediatric eso/rectal probe | Reusable, YSI-400 |
| MR420B adapter cable | Temperature adapter (2-pin) |
YSI-400 probes are interchangeable across BeneVision and BeneView. Confirm your standard with the YSI 400 vs YSI 700 compatibility test.
EtCO2 & water traps DryLine
All EtCO2| OEM cross-reference | Type |
|---|---|
| DryLine 115-043022-00 | Water trap (adult/pediatric) |
| DryLine 115-043023-00 | Water trap |
| DryLine 115-043024-00 | Water trap (neonatal) |
| DryLine 115-043025-00 | Water trap |
For mainstream / sidestream EtCO2 sensors and sampling lines, browse the EtCO2 collection.
Reading path by clinical scenario
Adult ICU (N15 / N17 / N19, fixed bedside)
5-lead 12-pin MPM ECG trunk plus grabber leadwires plus adult solid-gel electrodes; reusable adult finger-clip SpO2 (technology per host); adult-standard reusable NIBP cuff plus dual-tube extension; 1–2 IBP channels with a 12-pin Mindray-to-Edwards cable; YSI-400 skin temperature probe; EtCO2 if the MPM CO2 module is installed. Continuous monitoring with low patient turnover favours reusable accessories; closed-system pressure transducers reduce line entries.
Cardiac surgery OR (N22, sometimes N19)
10-lead diagnostic ECG trunk plus diagnostic leadwires plus radiolucent electrodes; reusable adult finger-clip or multi-site Y SpO2; adult-large or thigh NIBP cuff; 3–4 IBP channels with closed blood sampling; dual-site temperature; mainstream EtCO2 for intubated patients. Cardiac surgery needs simultaneous invasive monitoring across multiple sites, and esophageal core temperature is required for hypothermia protocols.
NICU (specialised N15 with neonatal module)
3-lead ECG trunk plus clip leadwires plus neonate radiolucent electrodes; neonate wrap reusable or neonate adhesive disposable SpO2; neonate NIBP sizes 1–5 with single-tube extension (confirm the A26 / A41 / A65 connector against the module); pediatric / neonate temperature probe; IBP umbilical artery monitoring if needed. Clinical reminder: 2-hourly site rotation and skin inspection are the primary defence against medical-device-related pressure injury (MDRPI) — see the Mindray neonate wrap SpO2 sensor page for detail.
Pediatric ICU (N12 / N15)
5-lead ECG trunk plus snap leadwires plus pediatric solid-gel electrodes; reusable pediatric finger-clip SpO2; pediatric NIBP cuff plus dual-tube extension; pediatric temperature probe; IBP and EtCO2 as needed.
Transport (N1)
N1 uses the 6-pin direct trunk: 5-lead 6-pin ECG trunk plus snap leadwires; SpO2 (7-pin Mindray-native is most common on the N1 — confirm); one or two NIBP cuff sizes; a strain-relief-reinforced SpO2 extension cable; a 6 m NIBP interconnect tube for the stretcher-to-monitor distance. Transport sees the highest mechanical stress on cables — the wide-grid dust-resistant strain relief addresses the connector-exit failure point.
PACU and step-down (N12 / N15)
5-lead ECG trunk plus snap leadwires plus adult solid-gel electrodes; reusable or disposable SpO2 (high turnover favours disposable); NIBP cuffs in 3–4 sizes per unit; optional temperature probe.
Mixed fleet — BeneVision N plus BeneView T
Hospitals running Mindray for 8+ years often have a mixed fleet: earlier BeneView T-series alongside current BeneVision N-series. What carries over:
| Accessory category | BeneView T ↔ BeneVision N | Notes |
|---|---|---|
| ECG leadwire sets | Interchangeable | Standard yoke-to-leadwire interface |
| ECG trunk cables | Generation-specific | T-series trunk pinout differs from the N MPM |
| ECG electrodes | Interchangeable | Universal |
| NIBP cuffs (Y-series) | Interchangeable | Same Mindray standard |
| SpO2 sensors | Depends on technology | Interchangeable only if both run the same SpO2 technology |
| Temperature probes (YSI-400) | Interchangeable | Same standard |
| IBP cables | Generation-specific | T-series and N-series use different IBP cabling |
When transitioning a fleet from BeneView T to BeneVision N over 12–24 months, standardise the interchangeable accessories (cuffs, electrodes, temperature probes, SpO2 where the technology matches) across both fleets and order generation-specific items (ECG trunks, IBP cables) only as needed per platform.
Common procurement pitfalls
- Ordering a 12-pin MPM trunk for the N1. The N1 uses the smaller 6-pin direct trunk. When buying ECG cables, specify "N1 transport" or "MPM module" by name.
- Ordering Mindray-native sensors when the monitor has Masimo SET or Nellcor OxiMax. Confirm the SpO2 technology per unit before ordering.
- Wrong neonate cuff connector (A26 vs A41 vs A65). The cuff body is identical; the connector varies. Photograph the existing cuff connector before reordering.
- Stocking reusable cuffs in only one size. A significant share of out-of-spec NIBP readings on adult patients trace to cuff-sizing mismatch rather than cuff brand. Stock 3–4 sizes per unit.
- Assuming BeneView T accessories transfer to BeneVision N. Some do, some do not — see the mixed-fleet table above.
- Ignoring the transducer brand on IBP adapter cables. A Mindray 12-pin IBP cable connects to an Edwards, Utah, B.Braun or Argon transducer depending on hospital protocol — the transducer-end fitting differs. Identify the disposable transducer in use first, then order the matching cable.
How to get a configuration review
Whether you run a pure BeneVision N fleet, a mixed N plus BeneView T fleet, or are planning expansion, MedLinket's configuration team will review your inventory. Provide your BeneVision and BeneView model counts, the SpO2 technology per model, the installed MPM module variants, your disposable transducer brand standard, and your current accessory inventory. There is no obligation to purchase.
Frequently asked questions
How do I tell which SpO2 technology my BeneVision has?
Check the SpO2 connector on the MPM module, the MPM module label, or the service menu (System Information → Module Status → SpO2 → Technology). Mindray-native uses a 7-pin connector, Masimo SET the LNCS connector, Nellcor OxiMax a 9-pin connector — each needs a different sensor family.
Will an N15 ECG trunk work on our N1 transport monitor?
No. The N15 uses a 12-pin MPM trunk; the N1 uses a smaller 6-pin direct trunk. They are not interchangeable. Stock both trunk types separately and specify the model by name on the purchase order.
Can BeneView T-series accessories be reused on BeneVision N?
Some — ECG leadwires, electrodes, NIBP cuffs and YSI-400 temperature probes carry over. ECG trunk cables and IBP cables do not, as the generations use different pinouts. SpO2 sensors carry over only if both platforms run the same SpO2 technology.
Compliance and quality. All MedLinket BeneVision-compatible accessories are manufactured under ISO 13485:2016 (TÜV), ISO 9001, FDA 510(k) clearance for U.S. market entry, CE marking, MDSAP and MHRA registration, with 100% per-unit electrical continuity QC. "Mindray," "BeneVision," "BeneView," "MPM," "Masimo," "SET," "LNCS," "Nellcor," "OxiMax," "Edwards," "Utah," "B.Braun" and "Argon" are trademarks of their respective owners. MedLinket is not affiliated with, endorsed by or licensed by Mindray Medical International or any other OEM. Use of OEM part numbers and model names is solely for compatibility identification, consistent with Magnuson-Moss Warranty Act protections.
About MedLinket. Founded 2004 in Shenzhen. NEEQ-listed (stock code 833505). Over 20 years specialising in patient-monitoring accessories, with extensive compatibility coverage for Mindray BeneVision and BeneView fleets. FDA 510(k), CE, MHRA, MDSAP, ISO 13485:2016 and ISO 9001 certified. Manufacturing across three production bases under Class 100,000 cleanroom standards. Serving 2,000+ hospitals across 117 countries and regions. Product liability insurance is carried across the product line, with cover up to USD 5 million; distributors may request a certificate of insurance within the policy period.