NIBP Readings: Systolic, Diastolic & MAP Ranges (Clinical Guide)

0 comments

 

📖 This article is part of our comprehensive Hospital Monitor Reading & Accessories Guide. Explore the full series for detailed guides on every vital sign parameter, alarm troubleshooting, and accessory selection.

By MedLinket Clinical Content Team | Updated: February 2026 | 15 min read

120/80
Normal adult reading (mmHg)
70–100
Normal MAP range (mmHg)
< 60
MAP danger threshold — act now

Quick Answer

Quick Answer

NIBP displays three numbers on a hospital monitor: Systolic (top number — pressure when the heart contracts), Diastolic (bottom number — pressure when the heart relaxes), and MAP (mean arterial pressure — average pressure during one cardiac cycle, shown in parentheses). Example: 120/80 (93). Normal adult blood pressure is approximately 120/80 mmHg with a MAP of 70–100 mmHg. A MAP below 60 mmHg signals inadequate organ perfusion and requires immediate clinical intervention.

What Does NIBP Mean on a Hospital Monitor?

NIBP stands for Non-Invasive Blood Pressure. It is one of the five core vital signs displayed on patient monitors in hospitals — alongside ECG, heart rate, SpO2, and EtCO2.

On most hospital monitors from Philips, GE Healthcare, Mindray, Draeger, and other manufacturers, the NIBP reading appears in a dedicated area of the screen — typically color-coded in red or white and labeled "NIBP" or "NBP." The display format is:

NIBP (mmHg)
120 / 80
(93)

Typical NIBP display on a hospital monitor: Systolic / Diastolic (MAP)

Unlike IBP (invasive blood pressure) monitoring — which requires an arterial catheter inserted directly into a blood vessel — NIBP uses an external inflatable blood pressure cuff wrapped around the patient's upper arm. This makes it the default method for blood pressure monitoring in general wards, emergency departments, pre-op areas, and many ICU settings.

The Three Numbers: Systolic, Diastolic, and MAP

Every NIBP reading on a hospital monitor gives you three distinct values. Here is exactly what each one represents and why it matters clinically.

Systolic Pressure (Top Number)

Systolic blood pressure (SBP) is the peak pressure exerted on the arterial walls when the left ventricle of the heart contracts during systole. It reflects the force of cardiac contraction and the elasticity of the arterial system.

  • Normal range for adults: <120 mmHg
  • Hypertension Stage 1: 130–139 mmHg
  • Hypertensive crisis: >180 mmHg — requires immediate evaluation

Clinical insight: Systolic pressure tends to rise with age as arteries stiffen. In neonates, a normal SBP may be as low as 60–80 mmHg (based on MedLinket internal training data for neonatal monitoring, where SBP 70–90 mmHg is the accepted reference range), which would be critically low for an adult patient. Always reference age-appropriate ranges.

Diastolic Pressure (Bottom Number)

Diastolic blood pressure (DBP) measures the residual pressure in the arteries when the heart relaxes between beats (diastole). It reflects the tone and resistance of the peripheral vascular system.

  • Normal range for adults: <80 mmHg
  • Hypertension Stage 1: 80–89 mmHg
  • Hypertensive crisis: >120 mmHg

Clinical insight: Isolated diastolic hypertension (high DBP with normal SBP) is more common in younger adults and may signal early vascular disease.

MAP — Mean Arterial Pressure (Number in Parentheses)

MAP (Mean Arterial Pressure) represents the average arterial blood pressure during one complete cardiac cycle. It is displayed in parentheses on the monitor — for example, (93).

The standard formula is:

MAP = DBP + 1/3 × (SBP − DBP)

Example: SBP 120, DBP 80 → MAP = 80 + 1/3 × (120 − 80) = 80 + 13.3 ≈ 93 mmHg

Why the formula weights diastole more heavily: At a normal resting heart rate, the heart spends roughly two-thirds of each cycle in diastole and only one-third in systole. MAP accounts for this timing asymmetry, making it a more accurate reflection of the true perfusion pressure reaching your organs.

🔑 Key Clinical Fact

In NIBP oscillometric monitoring, MAP is actually the only value directly measured by the monitor's algorithm. The algorithm identifies the point of maximal oscillation amplitude in the cuff as MAP, then uses proprietary calculations to derive the systolic and diastolic values. This makes MAP the most reliable number in your NIBP reading — a fact that experienced biomedical engineers frequently emphasize when training clinical staff.

Normal Blood Pressure Ranges by Population

Blood pressure targets differ significantly across patient populations. The following table summarizes the key reference ranges used in clinical practice.

Population Systolic (mmHg) Diastolic (mmHg) MAP (mmHg)
Healthy Adult <120 <80 70–100
Elderly (>65 yrs) <130* <80 70–105
Child (1–10 yrs) 80–110 50–75 55–85
Infant (1–12 months) 75–100 50–70 55–80
Neonate 60–90 30–60 40–60
ICU / Sepsis Target ≥65

← swipe table horizontally →

*ACC/AHA guidelines classify ≥130/80 as Stage 1 hypertension for adults. Elderly targets may be individualized by the treating physician.

Blood Pressure Categories (AHA/ACC Classification)

Category Systolic (mmHg) Diastolic (mmHg)
✅ Normal <120 <80
⚠️ Elevated 120–129 <80
⚠️ Hypertension Stage 1 130–139 80–89
🔴 Hypertension Stage 2 ≥140 ≥90
🚨 Hypertensive Crisis >180 >120

← swipe table horizontally →

When a blood pressure reading triggers an alarm on your monitor, see our detailed Blood Pressure Alarm Troubleshooting Guide for a step-by-step clinical response framework.

Why MAP Matters in Critical Care

While systolic and diastolic values are important for diagnosis, MAP is the number that critical care clinicians watch most closely — and for good reason. MAP reflects the actual driving pressure that perfuses vital organs, including the brain, kidneys, and heart.

MAP Range (mmHg) Clinical Significance
70–100 Normal — adequate organ perfusion
≥65 Minimum target in sepsis (Surviving Sepsis Campaign Guidelines)
<60 Organs begin to suffer ischemia; immediate intervention required
<50 Risk of irreversible organ damage; neuronal death may ensue
>110 Elevated — may indicate hypertension; increases vascular damage risk

← swipe table horizontally →

💡 Important Nuance From Clinical Practice

As experienced ICU nurses on clinical forums frequently discuss, MAP targets should be individualized. A patient with baseline hypertension may need a higher MAP (80–85 mmHg) to maintain adequate cerebral and renal perfusion than the standard 65 mmHg threshold. Always correlate MAP readings with clinical signs of perfusion — urine output, mental status, lactate levels, and capillary refill.

Understanding how alarm thresholds relate to these MAP values is critical. For a broader overview of all patient monitor alarm types and priority levels, read our False Alarms on Patient Monitors: Causes and Prevention guide.

↑ Back to top

How NIBP Measurement Works (Oscillometric Method)

Understanding how your patient monitor arrives at those three numbers helps you troubleshoot inaccurate readings and explain the process to patients and colleagues.

Hospital monitors measure NIBP using the oscillometric method. Based on MedLinket's internal product training materials, here is the process step by step:

  1. Inflation: The NIBP cuff inflates automatically to a pressure above the expected systolic level, temporarily occluding the brachial artery.
  2. Controlled deflation: The monitor gradually releases air from the cuff in controlled steps.
  3. Oscillation detection: As blood begins to flow through the partially compressed artery, tiny oscillations in cuff pressure are detected by a pressure sensor inside the monitor.
  4. MAP identification: The point at which oscillation amplitude is greatest corresponds to the mean arterial pressure (MAP) — this is the only value directly measured.
  5. SBP & DBP derivation: The monitor's proprietary algorithm then calculates systolic and diastolic values from the oscillation envelope. Each manufacturer (Philips, GE, Mindray, Draeger, etc.) uses a slightly different algorithm (where SP=MP/0.55 and DP=MP×0.85 mathematically in baseline models), which is why readings from different monitors may vary by a few mmHg. Note that oscillometric algorithms become unreliable when heart rate is <40 or >240 bpm, or SBP is >240 or <40 mmHg.

This is why experienced biomedical engineers often recommend paying closest attention to MAP — it is the directly measured value, while SBP and DBP are calculated estimates.

7 Factors That Affect NIBP Accuracy

Accurate blood pressure measurement depends on controlling multiple variables. Research shows that a single measurement mistake can skew readings by as much as 50 mmHg, potentially leading to misdiagnosis and inappropriate treatment. Here are the seven most common sources of error:

# Factor Impact on Reading Solution
1 Wrong cuff size Too small/narrow → falsely high (10–15 mmHg, up to +20 mmHg); too wide → falsely low Measure mid-arm circumference & use 3-step rule
2 Cuff over clothing Can add 5–50 mmHg to reading Place cuff on bare skin (<2 mm thin fabric only per protocol)
3 Cuff position relative to heart Every 10 cm above/below heart level ≈ 8 mmHg error Position cuff at heart level (4th intercostal space)
4 Patient movement / talking Talking adds 10+ mmHg; movement causes measurement failure Ensure patient is resting quietly for 3–5 mins
5 Full bladder Can raise SBP by up to 33 mmHg Allow patient to void prior to measurement
6 Arrhythmias / Low Perfusion Irregular rhythms or weak pulse amplitude disrupt oscillation detection May need manual auscultation or IBP monitoring
7 Worn or leaking cuff/hose Air leaks prevent proper inflation → "Cuff Error" / "Air Leak" Replace worn cuff (69.1% contaminated) & inspect NIBP hose

← swipe table horizontally →

💡 Pro Tip From Clinical Practice

When you get a reading that seems inconsistent with the patient's clinical picture, always repeat the measurement after verifying cuff size and placement. As many bedside nurses share on clinical forums: "If the number doesn't match the patient, believe the patient — not the monitor." A single NIBP reading should never be the sole basis for critical treatment decisions.

Choosing the Right Blood Pressure Cuff Size

Cuff sizing is the single most common source of NIBP measurement error in hospital wards and ICUs. Clinical studies and MedLinket engineering tests confirm that using a cuff that is too small or narrow forces the monitor pump to over-inflate to occlude the artery, artificially elevating systolic and diastolic blood pressure readings by 10–15 mmHg (and up to 20 mmHg) — frequently leading to misdiagnosis and unnecessary antihypertensive administration.

⚠️ Technical Warning: Cuff Size Mismatch Deviates NIBP Readings by 10–15 mmHg

Oscillometric Mechanics: According to MedLinket’s NIBP engineering protocols, applying an undersized or overly narrow cuff requires higher inflation pressure to compress the brachial artery, resulting in falsely elevated blood pressure readings by 10–15 mmHg. Conversely, an oversized or overly wide cuff distributes pressure across a broader surface, yielding falsely low readings that may mask true shock or hypotension.

MedLinket’s 3-Step Anti-Error Sizing SOP:

  • Step 1 (Mid-Arm Measurement): Use a flexible tape to measure upper arm circumference at the midpoint between the acromion (shoulder) and olecranon (elbow).
  • Step 2 (The 75% Bladder Rule): Verify that the cuff’s inflatable TPU bladder length covers at least 75% (AHA/MedLinket standard) of the patient's arm circumference.
  • Step 3 (Boundary Standard): If the measured circumference falls exactly on a size boundary (e.g., 32 cm between Adult and Large Adult), always default to the larger cuff size to prevent artificial pressure spikes.

For delicate NICU care, MedLinket’s Hylink TPU transparent neonatal series (#1–#5, 3–15 cm range) incorporates Index Lines and △ Artery Markers for rapid visual verification, allowing clinicians to assess skin condition directly while avoiding size-induced deviation. All MedLinket cuffs are 100% latex-free and DEHP-free.

To eliminate selection errors, MedLinket manufactures an 11-scale arm circumference matrix spanning from 3 cm (Premature Neonate #1) to 50 cm (Adult Thigh). Use this reference matrix for clinical purchasing and ward setup:

MedLinket Cuff Size Arm Circumference (cm) Patient Category & Clinical Application
Neonatal #1 3–6 Premature infants / NICU (Hylink TPU Transparent)
Neonatal #2–#5 4–15 Full-term neonate to infant (#2: 4-8cm, #3: 6-11cm, #4: 7-14cm, #5: 8-15cm)
Child 15–22 Pediatric wards & clinics
Small Adult 17–25 Slender adult arms / adolescent
Adult (Standard) 24–32 Average adult ward & outpatient (Universal)
Adult Long 28–37 Longer adult arm profile
Large Adult 32–42 Muscular or obese adult upper arm
Thigh / Bariatric 42–50 Bariatric patients / thigh measurement site

← swipe table horizontally →

Reusable adult NIBP cuff, TPU, Philips/GE/Mindray compatible, 27-35cm
Save 39% · Reusable TPU Cuff
Reusable NIBP Cuff — Adult (Universal, without Connector)
Soft TPU bladder cuff · 27–35 cm range · latex-free & DEHP-free · compatible with Philips, GE, Mindray, Draeger
$9.20 USD $15.00
View Product →

For a detailed breakdown of ward-wide cuff standardization, read our dedicated procurement guides:

↑ Back to top

When to Escalate: NIBP vs IBP Monitoring

NIBP is adequate for most ward and outpatient scenarios, but in critical care, there are situations where invasive blood pressure (IBP) monitoring via an arterial line becomes necessary. Based on established clinical guidelines and MedLinket's internal patient monitoring training materials, consider transitioning from NIBP to IBP cables and IBP transducers when:

  • Systolic BP >250 mmHg or <50 mmHg (beyond reliable NIBP oscillometric range)
  • Rapid BP fluctuations >40 mmHg within minutes requiring real-time wave analysis
  • Major cardiac surgery, organ transplant, or high-risk neurosurgical procedures
  • High-dose vasopressor therapy (where peripheral vasoconstriction skews NIBP accuracy)
  • Repeated arterial blood gas (ABG) sampling is required (closed blood sampling system needed)
  • Persistent NIBP measurement failures due to severe arrhythmias, tremor, or shock

For hospitals that need IBP monitoring accessories, MedLinket offers compatible IBP cables and NMPA/FDA-cleared disposable IBP transducers (Registration No. 20243072401). Featuring US-imported piezoresistive silicon chips with ±2% sensitivity, integrated 3ml/h flush valves, and closed-loop blood sampling (R-series with PTFE bacteria filter), MedLinket IBP systems reduce CLABSI infection risks while maintaining continuous arterial waveform fidelity.

NIBP Troubleshooting: When "Measurement Failed" Appears

One of the most common frustrations at the bedside is the dreaded "NIBP Measurement Failed" or "Cuff Error" alarm. Here is a systematic approach to resolve it quickly.

Problem Alarm Likely Cause Step-by-Step Solution
Measurement failed — no reading Patient movement; severe tremor; arrhythmia; loose cuff velcro Settle patient, re-align △ marker over brachial artery, retry
"Cuff error" / "Air leak" alarm Damaged TPU bladder; loose NIBP connector (A05/A41/A26); kinked hose Check connector seating; replace worn NIBP hose or cuff
Abnormally high/low reading Wrong cuff size (small = +10-15mmHg); arm not at heart level Apply 3-Step Sizing SOP; level arm to 4th intercostal space
Inconsistent readings (vary >15 mmHg) Pain; anxiety; white-coat effect; rapid cycle inflation Allow 2-min rest between cycles; average 3 readings
Cuff inflates but never deflates Bleed valve failure inside monitor NIBP module Remove cuff immediately to avoid ischemia; call Biomed engineering
Repeated failures in obese/conical arms Standard rectangular cuff slipping down conical arm Use Large Adult / Thigh cuff (42–50 cm), or place cuff on forearm

← swipe table horizontally →

If the cuff fails to inflate at all from the start, check our dedicated walkthrough on what to check when the NIBP cuff won't inflate at all. For general equipment issues, see Patient Monitor No Reading Troubleshooting Guide.

Philips compatible NIBP hose 989803209771, TPU jacket, quick-connect, 10ft
Save 20% · TPU Jacket · Philips Compatible
Philips-Compatible NIBP Hose (Quick-Connect) — 989803209771
TPU-jacket air hose · 10 ft (3 m) · single tube · adult/pediatric · for Philips IntelliVue / MP / MX / Goldway series
$40.00 USD $50.00
View Product →

↑ Back to top

Proper NIBP Cuff Application: 8-Step Clinical SOP

Correct cuff application is the foundation of accurate NIBP measurement. Based on MedLinket's clinical training protocol, follow this 8-step standard operating procedure:

  1. Confirm patient category — adult, pediatric, or neonatal — to select the correct cuff group.
  2. Measure mid-upper arm circumference with a flexible tape. On size boundaries, select the larger cuff.
  3. Ensure the cuff bladder is fully deflated before wrapping around the limb.
  4. Remove thick clothing — ensure maximum 2 mm fabric between cuff and skin per protocol.
  5. Align the △ (Artery Marker) printed on the cuff directly over the brachial artery.
  6. Wrap snugly — verify fit using the 2-finger rule (two fingers inserted beneath cuff without force).
  7. Position the arm at heart level (4th intercostal space), supported, with approximately 45° abduction.
  8. Verify placement — check that the cuff is not twisted, kinked, or overlapping the antecubital fossa.
⚠️ Contraindicated Limbs — Do NOT Apply NIBP Cuff On:
  • Same arm as an active IV infusion line or arterial catheter
  • Arm with severe thrombosis, lymphedema, or AV fistula (dialysis access)
  • Arm with burns, trauma, skin grafts, or active lesions
  • Ipsilateral arm following mastectomy with lymph node dissection

Accurate NIBP readings depend not only on technique but also on the quality and compatibility of your monitoring accessories. Studies reveal that over 69.1% of reusable NIBP cuffs harbor bacterial contamination (including MRSA), making single-patient disposable cuffs essential for isolation wards, burn units, and ICUs.

Accessory Category Function & Material Lifespan / Ward Use MedLinket Solution
Reusable NIBP Cuff Nylon + TPU bladder; washable & latex-free 6–12 months Adult Reusable TPU Cuff
Disposable NIBP Cuff Soft non-woven fabric or Yellow Isolation (Y006D) Single patient use Disposable Adult Long Cuff
Neonatal Hylink Cuff Transparent TPU material for visual skin check NICU single patient Hylink TPU Neonate #3 (M1870A)
ABPM Cuff 24-hour ambulatory blood pressure monitoring 6–12 months ABPM Cuff with Bayonet Connector
NIBP Air Hose TPU jacketed single/dual lumen hose (1.5–3m) 2–3 years Philips NIBP Hose Guide
NIBP Connectors Precision connectors (A05/A26/A41/A57/A108/A118) Matches hose lifespan All NIBP Connectors Matrix

← swipe table horizontally →

Build a Complete Ward NIBP Accessory Fleet

Standardize your ward, NICU, and patient monitors with OEM-compatible, certified NIBP parts

Adult reusable NIBP cuff TPU
Adult Reusable Cuff
Soft TPU · 27–35 cm
$9.20
Save 39%
View Product →
Neonate TPU NIBP cuff M1870A
Neonate #3 TPU Cuff
M1870A · Box of 24
$44.00
Save 17%
View Product →
Philips NIBP hose quick-connect
Philips NIBP Hose
989803209771 (10 ft)
$40.00
Save 20%
View Product →

Mix reusable, disposable, and monitor-side hose assemblies to equip general wards, NICU, and ICU beds.

For a complete overview of all patient monitor accessories organized by parameter type — including SpO2 sensors, ECG cables, and temperature probes — see our dedicated guide. Need help determining compatibility? Read How to Identify Which Cables Your Monitor Needs or learn the differences between OEM vs Compatible Accessories. If you're evaluating suppliers, understand what "compatible" means for medical accessories beyond fitting a plug.

Why Hospitals in 117+ Countries Trust MedLinket NIBP Accessories

MedLinket (Shenzhen Med-Link Electronics Tech Co., Ltd) has specialized in medical sensors, cables, and monitoring consumables since 2004. Listed on China's NEEQ exchange (stock code: 833505), MedLinket operates a 10,000+ m² manufacturing footprint with 2,800+ sets of precision molds, supplying 2,000+ hospital end-users worldwide.

Certifications FDA 510(k) · CE (MDR) · ISO 13485:2016 · MDSAP · NMPA (Registration No. 20243072401)
Compatible Brands Philips, GE Healthcare, Mindray, Draeger, Masimo, Nihon Kohden, ZOLL, Colin, Welch Allyn, and 30+ more
Global Reach Exported to 117+ countries · 2,000+ hospital end-users · 14 regional markets with annual sales >$1M USD
Product Liability Insurance $5 Million USD global coverage (Liberty Insurance) — additional insured certificates available for distributors
Manufacturing Capacity 2 owned factories (Shenzhen & Shaoguan) · 2,822 sets of molds · 10,303 product styles · 100% factory inspection

← swipe table horizontally →

↑ Back to top

This article is part of our comprehensive Hospital Monitor Reading & Accessories Guide. Explore other articles in this cluster series:

← swipe table horizontally →

Frequently Asked Questions

What is a normal NIBP reading on a hospital monitor?

A normal NIBP reading for a healthy adult is below 120/80 mmHg with a MAP of 70–100 mmHg. The monitor displays three numbers: systolic/diastolic (MAP). For pediatric and neonatal patients, normal values are significantly lower — see the population-specific ranges table above.

What does MAP mean on a hospital monitor?

MAP (Mean Arterial Pressure) is the average arterial pressure during one cardiac cycle. It is the best single indicator of organ perfusion pressure. Calculated as DBP + 1/3 × (SBP − DBP). A MAP below 60 mmHg means vital organs are not receiving adequate blood flow. The Surviving Sepsis Campaign targets a minimum MAP of 65 mmHg in septic shock patients.

How much can an incorrect cuff size skew my blood pressure reading?

According to MedLinket engineering and clinical protocols, using a cuff that is too small or narrow forces the pump to over-inflate, causing falsely elevated readings of 10–15 mmHg (and up to 20 mmHg). Conversely, an oversized cuff yields falsely lower values. Always measure upper arm circumference at the midpoint and ensure the inflatable TPU bladder covers at least 75% of the arm circumference. On size boundaries, default to the larger size.

Why do I get different NIBP readings each time?

Blood pressure naturally fluctuates with respiration, pain, activity, and body position. Differences of 5–10 mmHg between consecutive readings are normal. Larger variations often stem from incorrect cuff size, patient movement, arm position above or below heart level, or a leaking NIBP hose. Always allow a 1–2 minute rest interval between repeated measurements.

What does it mean when NIBP measurement fails?

A failed measurement means the monitor's pressure sensor could not detect reliable pulse oscillations. Common causes include patient movement, severe tremor, arrhythmias, loose cuff velcro, air leaks in the bladder, or cuff placed over thick clothing. See the troubleshooting table above for step-by-step solutions. For persistent failures with irregular rhythms, consider manual auscultation or IBP monitoring.

What is the difference between NIBP and IBP?

NIBP uses an external inflatable cuff on the arm and the oscillometric method — non-invasive, intermittent, and suitable for general wards and stable patients. IBP uses an arterial catheter inserted directly into a blood vessel for continuous, real-time waveform readings. IBP is required for critically ill ICU patients, those on high-dose vasopressors, or when SBP exceeds 250 or drops below 50 mmHg.

How often should NIBP cuffs be replaced?

Reusable NIBP cuffs last 6–12 months in high-use hospital wards. Replace when the TPU bladder leaks, fabric frays, or velcro fails. Disposable NIBP cuffs are single-patient use. Studies show over 69.1% of reusable cuffs harbor pathogenic bacteria, making single-patient disposable cuffs mandatory in isolation wards, burn units, and ICUs. See our Accessory Replacement Schedule for detailed intervals.


Need Help Finding the Right NIBP Cuff or Hose for Your Monitor Fleet?

Tell us your monitor brand and model (Philips, GE, Mindray, Draeger, Nihon Kohden, etc.). Our biomedical engineers will verify connector compatibility and send you a matched product recommendation and bulk quote — free of charge.

📧 Email Us (shopify@medlinket.com) 💬 WhatsApp Us

MOQ: Low wholesale minimums · Lead Time: 3–5 business days · FDA 510(k) & CE MDR Certified


About This Article
Published by the MedLinket Clinical Content Team. Our team includes biomedical engineers and clinical application specialists with hands-on experience across Philips, GE, Mindray, and Draeger monitoring systems. Content is reviewed against MedLinket's internal product training manuals, ISO 13485 quality standards, and AHA/AAMI clinical guidelines.

Medical Disclaimer: This article is intended for educational purposes for healthcare professionals, biomedical engineers, and procurement officers. It does not constitute individual medical advice. Always follow your facility's clinical protocols and consult with qualified physicians for patient treatment decisions.

Last Updated: February 2026 · Review Cycle: Every 6 months


ECG Waveform & ST Segment Guide: Reading Hospital Monitor Numbers

What is EtCO2 and Why is It Monitored? | Normal Ranges & Clinical Guide

Leave a comment

Please note, comments need to be approved before they are published.

Declaration:

  • All other companies and brand names mentioned on this page are for identification purposes only and do not imply any affiliation, partnership, or endorsement of our products
  • The picture and the object differ slightly in appearance (e.g., connector design, color), but function the same.